Accurate calibration of a 1000N tension force sensor ensures reliable measurements in industrial, aerospace, and research applications. This guide provides a step-by-step methodology, advanced techniques, and best practices to maintain precision. Below, we explore calibration procedures, equipment requirements, and troubleshooting strategies while incorporating visual aids and practical examples.
A 1000N tension force sensor measures pulling forces up to 1,000 Newtons. These sensors are critical in applications like material testing, conveyor systems, and structural monitoring. Calibration aligns the sensor's output with known force standards, compensating for mechanical wear, environmental changes, or electrical drift[4][9].
Key Components:
- Load cell (strain gauge or piezoelectric)
- Signal amplifier
- Data acquisition system
- Calibration weights or force standards
A. Equipment and Environment
- Calibration Weights: Use certified weights covering 10–100% of the 1000N range (e.g., 100N, 500N, 1000N)[12][14].
- Stable Environment: Control temperature (±1°C) and humidity (40–60% RH) to minimize thermal expansion and electrical noise[4][8].
- Mounting Hardware: Ensure adapters and fixtures match the sensor's threading (e.g., M10 eyebolts) to prevent misalignment[1][10].
B. Initial Checks
- Inspect for physical damage or loose connections[3][10].
- Verify zero balance (sensor output with no load)[4][8].
- Confirm amplifier settings (e.g., 10V output range)[6][13].
Step 1: Zero Calibration
- Remove all external loads.
- Adjust the sensor's output to 0N using the amplifier's zeroing function[6][11].
- Validate stability over 5 minutes (drift <0.1% FS)[8][14].
Step 2: Apply Known Loads
Load (N) | Expected Output (V) | Tolerance (±%) |
---|---|---|
100 | 1.0 | 0.5 |
500 | 5.0 | 0.3 |
1000 | 10.0 | 0.2 |
- Suspend weights using calibrated pulleys or direct threading[2][5].
- Record sensor output at each load[9][13].
Step 3: Linear Regression Adjustment
- Plot applied force vs. sensor output.
- Calculate slope (k) and intercept (b) using:
F=k⋅V+b
- Update amplifier parameters to match the equation[7][13].
Step 4: Hysteresis Testing
- Cycle loads ascending (0→1000N) and descending (1000N→0).
- Acceptable hysteresis: <0.25% FS[4][9].
Step 5: Repeatability Verification
- Apply 500N three times.
- Maximum deviation: <0.15% FS[12][14].
- Temperature Compensation: Calibrate at 20°C, 40°C, and -10°C to model thermal drift[9][14].
- Dynamic Calibration: Use servo-controlled actuators to simulate real-world force rates (e.g., 50N/s)[9].
- Multi-Axis Calibration: For 2-axis sensors, apply orthogonal forces and adjust cross-sensitivity coefficients[3][9].
Error Type | Cause | Solution |
---|---|---|
Non-linear output | Misaligned adapters | Use alignment plugs[1][10] |
Zero drift | Temperature fluctuations | Re-calibrate in stable env[4] |
Overload damage | Exceeding 1300N limit | Replace load cell |
- Record calibration date, weights used, environmental conditions, and adjustment values[8][9].
- Attach a calibration certificate traceable to ISO/IEC 17025 standards[12][14].
Calibrating a 1000N tension force sensor requires meticulous preparation, systematic load application, and rigorous validation. By adhering to standardized procedures and leveraging advanced techniques like temperature compensation, users can achieve ±0.2% FS accuracy. Regular maintenance and documentation further ensure long-term reliability in critical applications.
Calibrate annually or after exposure to shocks, overloads, or extreme temperatures[4][12][14].
No—certified calibration weights with ±0.1% accuracy are mandatory[12][14].
Internal mechanical stops or adapter bottoming-out may limit the range—inspect and replace components[13].
Temperature changes >5°C can induce ±0.5% FS drift; always calibrate in controlled conditions[4][8][9].
Yes, but perform axis-specific calibrations and adjust cross-talk coefficients[3][9]
[1] https://mhforce.com/wp-content/uploads/2021/05/Force-Calibration-Guidance-for-Beginners.pdf
[2] https://www.fms-technology.com/fileadmin/import/amplifier-product_group/operating_instructions_EMGZ309_EN.pdf
[3] https://www.xjcsensor.com/how-to-maintain-and-calibrate-2-axis-force-sensors/
[4] https://www.fibossensor.com/how-to-identify-calibration-errors-in-compression-and-tension-load-cells.html
[5] https://haehne.de/fileadmin/erklaerungen_zertifikate/practice_guide.pdf
[6] https://www.youtube.com/watch?v=_8jPupKtD0M
[7] https://www.tekscan.com/blog/flexiforce/video-how-calibrate-and-adjust-sensitivity-embedded-flexiforce-sensor
[8] https://tacunasystems.com/knowledge-base/calibrating-the-force-measuring-system/
[9] https://www.fibossensor.com/how-to-calibrate-a-force-tension-sensor-for-accurate-readings.html
[10] https://www.arisewebguiding.com/how-to-maintain-and-calibrate-tension-control-systems-effectively
[11] https://www.vernier.com/manuals/vsmt-force/
[12] https://www.alluris.com/calibration/80.ekan
[13] https://www.vernier.com/files/manuals/vsmt-force/vsmt-force.pdf
[14] https://www.pce-instruments.com/english/measuring-instruments/test-meters/tension-dynamometer-tension-load-cell-pce-instruments-tension-dynamometer-tension-load-cell-pce-dfg-1k-x-ica-incl.-iso-calibration-certificate-det_6211245.htm?_list=qr.art&_listpos=302
[15] https://www.vernier.com/files/manuals/vsmt-force/vsmt-force.pdf
[16] https://www.hans-schmidt.com/media/user-manual-SFZ-SFK-SFD-E.pdf
[17] https://www.nordela.se/ws_media/pdf/FMS%20EMGZ%20310_e_Manual.pdf
[18] https://www.arisewebguiding.com/how-to-maintain-and-calibrate-tension-control-systems-effectively
[19] https://www.hans-schmidt.com/media/user-manual-ZT-E.pdf
[20] https://www.sciencedirect.com/science/article/pii/S266591742400312X
[21] https://www.hans-schmidt.com/en/produkt-details/force-gauge-rdsv-1000n/
[22] https://help.mecmesin.com/docs/advanced-force-gauge-afg-operating-manual-includes-afti-display
[23] https://www.fibossensor.com/how-to-calibrate-a-force-tension-sensor-for-accurate-readings.html
[24] https://mhforce.com/compression-and-tension-forces-calibration/
[25] https://www.ahlborn.com/assets/uploads/general/Produkte/Sensoren/Zug-und-Druckkraft-Sensor/Force-english.pdf
[26] https://www.lorenz-messtechnik.de/phplogin/login_en/html/kraft.php
[27] https://www.checkline.com/product/FS06/FS06-200
[28] https://www.checkline.com/product/R07/MR07-200
[29] https://www.vernier.com/manuals/vsmt-force/
[30] https://www.hbm.com/es/10688/the-load-cell-calibration-standard-iso-376/
[31] https://www.youtube.com/watch?v=HX7UipDhkUk
[32] https://mrmmetrology.com/product/mr01-200-mark-10-force-sensor/
[33] https://www.adinstruments.com/support/videos/how-calibrate-force-transducer-and-apply-pre-tension
[34] https://www.wika.com/en-in/tension_compression_force_transducers.WIKA
[35] https://www.hbm.com/it/10688/the-load-cell-calibration-standard-iso-376/
[36] https://www.octogon.org/en/product/force-transducer-force-sensors/tension-compression-force-sensors
[37] https://haehne.de/en/service/faq/calculations-sensor-and-amplifier-signals
[38] https://haehne.de/en/web-tension/tension-force-sensor/tension-force-sensor-zak
[39] https://mhforce.com/tension-link-calibration/
[40] https://www.pce-instruments.com/english/measuring-instruments/test-meters/tension-dynamometer-tension-load-cell-kat_162467.htm
[41] https://www.fibossensor.com/how-to-identify-calibration-errors-in-compression-and-tension-load-cells.html
[42] https://electronics.stackexchange.com/questions/312460/measuring-tension-force-on-rope
[43] https://www.strainsert.com/faq-on-force-sensor-calibration/
[44] https://haehne.de/en/web-tension/tension-force-sensor/tension-force-sensor-zad
[45] https://www.futek.com/calibration-services-FAQ
[46] https://haehne.de/en/web-tension/tension-force-sensor
[47] https://www.andilog.com/wlc-slc-wireless-bluetooth-load-cell-force-sensor.html
[48] https://www.fms-technology.com/fileadmin/import/amplifier-product_group/operating_instructions_EMGZ310_ComACT_EN.pdf
[49] https://ras.papercept.net/images/temp/AIM/files/0073.pdf
[50] https://mark-10.com/products/indicators-sensors/force-sensors/r01/
content is empty!
Contact:
Phone: +86 18921011531
Email: nickinfo@fibos.cn
Add: 12-1 Xinhui Road, Fengshu Industrial Park, Changzhou, China