FIBOS has completed two high-capacity force calibrations:
FA4249: calibration completed at the 400-tonne level.
FA4250: calibration completed at the 700-tonne level.
For engineers who design, build, or verify prestressing equipment, these two models are concrete starting points when a project needs force measurement in the 400- or 700-tonne class.
This article explains how a force calibrator supports prestressing jack verification, what the stated calibration levels confirm, and what to check before integrating either model into a test system.

Large force calibration equipment with an adjacent lifting gantry.
Prestressing equipment uses hydraulic jacks to apply force to steel tendons or bars. The jack shows hydraulic pressure, but the project depends on the force actually delivered.
A pressure reading alone does not establish that force for every jack and operating condition. Engineers therefore need a measured force reference to relate the two.
At 400 or 700 tonnes, the measuring element is only part of the challenge. A reliable setup also depends on:
A reaction structure designed for the test load.
Loading surfaces and alignment that keep the force on the intended axis.
Force measurement that covers the planned working points with suitable uncertainty and usable resolution.
A force calibrator provides a measured force reference inside a properly designed loading arrangement. A typical verification follows three steps:
Place the force calibrator in the jack's load path, within a reaction frame designed for the test load.
Apply load in steps and record the jack's pressure indication and the measured force at each selected point.
Use the paired readings to build a pressure-to-force relationship for the jack and gauge combination that was tested.
The FHWA Post-Tensioning Tendon Installation and Grouting Manual describes this use of load cells in jack calibration and notes that the resulting chart applies to the particular jack and gauge combination tested.
Keep two activities separate: calibrating the force-measuring instrument, and verifying the complete jack system. The assembled system still needs defined fixtures, readout equipment, loading procedure, and acceptance criteria.

Loading platen and working space inside the force calibration frame.
Choose the model by the forces your project actually needs to measure, not by the headline tonnage alone.
Calibration levels confirmed by FIBOS
| Model | Completed calibration level | Project assessment focus |
|---|---|---|
| FA4249 | 400 tonnes | Working points, documented measurement performance, and mechanical fit |
| FA4250 | 700 tonnes | Working points, load-path capacity, and mechanical fit |
The FA4249 suits discussions about force verification in the 400-tonne load class. Start with the minimum and maximum forces you will measure.
Then check whether the documented calibration points, measurement uncertainty, and readout resolution suit those working points.
The FA4250 brings a higher calibration level to projects involving larger-force prestressing equipment and test arrangements.
A higher force level also means reviewing the whole load path. The frame, adapters, contact surfaces, and handling arrangements each need their own engineering assessment.
For projects requiring force measurement above the 700-tonne class, FIBOS has also completed compression calibration of its FA4221-10MN force sensor, with recorded test forces reaching 10,000 kN (10 MN). Read our 1,000-tonne-class force sensor calibration article for the test points, results, and integration considerations when evaluating higher-force measurement.
Reading the stated calibration level
| Item | Status |
|---|---|
| Model and completed calibration level | Confirmed by FIBOS: FA4249 at 400 tonnes; FA4250 at 700 tonnes |
| Actual test points and force units | Check the calibration documentation |
| Uncertainty and resolution at your working points | Check the calibration documentation |
| Accuracy class and full calibrated interval | Not defined by the tonne figure alone |
| Permissible overload | Not implied by the calibration level |
| Suitability of your test rig and readout | Requires engineering assessment of the complete system |
A useful assessment starts with the equipment drawing and the measurement task. Prepare the following before choosing a model and configuration:
Integration checklist
| Check | What to define or provide |
|---|---|
| Force range | Lowest working point, normal operating forces, and maximum planned test force |
| Mechanical fit | Available height, contact diameters, mounting details, and clearance around the load path |
| Loading conditions | Force direction, centering method, load sequence, and holding times; assess eccentric and side loads |
| Measurement chain | Sensor output, excitation, cable, connector, and compatibility with the intended indicator or data acquisition system |
| Verification plan | Test points, repeat measurements, loading and unloading requirements, and acceptance limits, agreed before testing |
| Replacement (if applicable) | Existing model, installation drawing, output specification, and current calibration record |
When replacing an existing instrument, matching the nominal load class alone does not establish interchangeability.

Force calibration frame, control cabinet, and lifting equipment.
FIBOS has completed 400-tonne calibration for the FA4249 and 700-tonne calibration for the FA4250. For project specifications, use the force units and test points stated in the calibration documentation.
The relationship between pressure and delivered force has to be established for the specific jack and gauge combination. A measured force reference, such as a force calibrator, provides that link.
No. A headline calibration level does not establish performance at every lower load. Check the documented calibration points, uncertainty, and readout resolution against your own working points.
No. They are separate activities. The complete system also needs suitable fixtures, readout equipment, loading procedure, and acceptance criteria.
It may be evaluated, but matching the nominal load class is not enough. Provide the existing model, installation drawing, output specification, and calibration record for assessment.
FIBOS Completes 400- and 700-Tonne Calibration: Force Measurement for Prestressing Equipment
Piezoelectric Force Sensors: Types, Applications, and How to Choose
FIBOS Completes 10 MN Force Sensor Calibration: A Milestone in 1,000-Tonne-Class Measurement
Column Force Sensors: Types, Applications, and How to Choose
Top 10 Force Sensor Manufacturers for Buyers in Indonesia 2026
Top 15 Force Sensor Manufacturers for Buyers in Germany 2026
Inline Tension & Compression Force Sensors: Types, Applications, and How to Choose
Through-Hole & Donut Force Sensors: Types, Applications, and How to Choose
Multi-Axis Force Sensors: Types, Applications, and How to Choose
Precision Force Sensing
— Engineered for Stability, Calibrated for Accuracy.